COD Tester
COD Meter
COD Meter

COD Meter

Rapid digestion COD meter for chemical oxygen demand testing. Benchtop and portable options, 20 min digestion, wide measuring range, chloride interference suppression. Suitable for wastewater, surface water and environmental lab water quality analysis.

Product Introduction

Technical Parameters

Product Video

Procurement Guide

Product Overview


The COD Analyzer (Chemical Oxygen Demand Analyzer) complies with **HJ/T399-2007 “Water Quality—Determination of Chemical Oxygen Demand—Rapid Digestion Spectrophotometric Method”**, It employs the rapid digestion spectrophotometric method using potassium dichromate to measure the Chemical Oxygen Demand (COD) in water, reflecting the degree of organic pollution in the water. Available in both benchtop and portable models, it can be integrated with a digestion unit for streamlined testing. With simple operation and fast testing speeds, it replaces the traditional reflux method and is widely used in water quality testing scenarios such as environmental monitoring stations, wastewater treatment plants, industrial enterprises, and research institutions across the People’s Republic of China...


Working Principle


In a strongly acidic system, potassium dichromate oxidant and silver sulfate catalyst are added to the water sample, which is then rapidly digested under sealed conditions at 165°C. Organic matter in the water is oxidized, and hexavalent chromium is reduced to trivalent chromium; Changes in absorbance are measured using spectrophotometry, and the instrument automatically calculates the COD concentration. An accompanying masking agent suppresses chloride ion interference, ensuring testing accuracy. People’s Republic of China...


Key Features


1. **Rapid Testing**: Constant-temperature digestion at 165°C completes the digestion of a batch of samples in 20 minutes. The entire testing cycle is short, significantly saving time compared to the traditional reflux method.

2. **Wide, Segmented Measurement Range**: Freely switch between multiple ranges to test both low-concentration surface water and high-concentration industrial wastewater, accommodating various water samples.

3. **Tube-Based Colorimetric Design**: Testing is performed directly in pre-filled digestion tubes without the need to transfer samples to cuvettes, reducing human error and ensuring convenient operation.

4. **Intelligent Touchscreen Operation**: High-definition color touchscreen with built-in calibration curves; supports single-point and multi-point calibration, and includes a curve-saving function.

5. **Comprehensive Data Management**: Built-in thermal printer for instant printing of test reports; supports USB export and data storage, and can be integrated with platforms for data upload.

6. **Stable Optical System**: Imported cold light source ensures excellent optical stability, long light source lifespan, high repeatability, and low error rates.

7. **Portable Option**: The portable model features a built-in lithium-ion battery, suitable for field sampling and emergency monitoring; the desktop model is suitable for routine laboratory testing.

8. **Interference-Resistant Design**: Equipped with a chloride ion masking solution, it can withstand testing of high-chloride water samples and is suitable for textile printing and dyeing, chemical processing, and high-salinity wastewater.


Applications


- Online and laboratory COD testing of influent and effluent at wastewater treatment plants; discharge control

- Monitoring of surface water, groundwater, and wastewater at ecological and environmental monitoring stations and third-party testing agencies

- Industrial wastewater testing in sectors such as chemicals, dyeing and printing, papermaking, electroplating, food processing, and metallurgy

- Water quality-related experimental research at universities and research institutes

- Rapid monitoring of rivers, lakes, black and odorous water bodies, and emergency pollution sites


Summary of Product Advantages


The COD analyzer balances compliance with national standards and testing efficiency. It features integrated digestion and colorimetric analysis, with pre-filled reagent tubes that are ready to use out of the box, making it easy to get started. Test results are stable and reliable, and data can be saved and printed. The benchtop model is suitable for laboratories, while the portable model is ideal for outdoor emergency response, making it the mainstream instrument for routine COD testing in water quality analysis.


Translated with DeepL.com (free version)


Technical Specifications


Table


| Item | Parameter |

| --- | --- |

| Testing Standard | HJ/T399-2007 Rapid Digestion Spectrophotometric Method |

| Measurement Range | 0–15,000 mg/L (multi-range) |

| Detection Limit | 5 mg/L |

| Digestion Temperature | 165°C, PID temperature control |

| Digestion Time | 20 min |

| Indicated Value Error | ≤±5% |

| Repeatability | ≤3% |

| Optical Stability | ≤±0.003 Abs/20 min |

| Colorimetric Tube Specifications | Φ16 mm sealed digestion tubes |

| Data Output | USB, thermal printer, Wi-Fi (optional) |

| Power Supply | AC 220 V (desktop model); lithium-ion battery (portable model) |

| Operating Environment | 5–40 °C, humidity ≤ 85% RH |